What are the differences between measurement results obtained by image analysis and laser diffraction particle size analysis?

2023-07-17WIKI

The measurement result obtained by the laser diffraction particle size analysis is based on the scattered signals of particles captured onto detector arrays. An inverse algorithm is carried out to calculate a theoretical signal intensity distribution, which is compared to the measured one. An index called residual is used to describe the difference between the theoretical and measured signal intensity distributions. The theoretical signal intensity distribution with the smallest residual value is considered the measurement result. In contrast, the measurement result obtained by image analysis is calculated based on the captured images of real particle projections instead of algorithms.

 

The laser diffraction particle size analysis yields signal-equivalent diameters, which is suitable for the analysis of small micro-level particles, while image analysis provides area-equivalent diameters, measurement of oversized particles is therefore possible.

 

 

 

Download-Beginner-Guide-to-Particle-Analysis

Struggling to grasp particle analysis concepts? Our new Beginner's Guide to Particle Analysis is here to help you grasp these concepts in the most accessible manner possible. Download it as a PDF to read on the train, on the plane, or wherever else there’s no internet.

 

 

Rate this article

Share On

share
facebook twitter linkedin

Related Particle Size Analyzer

  • BeNano 180 Zeta Pro

    BeNano 180 Zeta Pro

    Nanoparticle Size and Zeta Potential Analyzer

    Technology: Dynamic Light Scattering, Electrophoretic Light Scattering, Static Light Scattering

  • Bettersizer S3 Plus

    Bettersizer S3 Plus 126

    Laser Diffraction Particle Size and Shape Analyzer

    0.01 - 3,500μm(Laser System)

    2 - 3,500μm(Image System)

  • Bettersizer ST

    Bettersizer ST

    One-stop Particle Size Analyzer

    Dispersion type: Wet

    Size range: 0.1 - 1,000µm

    Repeatability: ≤1% variation