How to choose a proper obscuration range?

2022-12-07WIKI

In general, laser diffraction particle size analysis is carried out when the obscuration of the sample in the flow stream is between 10 and 15%. A proper working obscuration lies between having too few particles thus affecting statistical representativeness, and too many particles leading to multiple scattering and hence undersizing of the particle results. In other words, if the obscuration is set between 10 and 15%, multiple scattering can be avoided, and sampling representativeness can be guaranteed.

 

Small particles have a more significant impact on the obscuration. For samples of the same volume, samples with smaller particle size distributions have more particles, leading to higher obscurations and, therefore, the potential for multiple scattering if the obscuration is higher than the upper recommended obscuration limit. According to experience, the proper obscurations for large particles are higher than those for smaller particles during measurement.

 

How the obscuration work both for few particles and too many particles

 

 

 

 

Download-Beginner-Guide-to-Particle-Analysis

Struggling to grasp particle analysis concepts? Our new Beginner's Guide to Particle Analysis is here to help you grasp these concepts in the most accessible manner possible. Download it as a PDF to read on the train, on the plane, or wherever else there’s no internet.

 

 

Rate this article

Share On

share
facebook twitter linkedin

Related Particle Size Analyzer

  • BeNano 180 Zeta Pro

    BeNano 180 Zeta Pro

    Nanoparticle Size and Zeta Potential Analyzer

    Technology: Dynamic Light Scattering, Electrophoretic Light Scattering, Static Light Scattering

  • Bettersizer S3 Plus

    Bettersizer S3 Plus 126

    Laser Diffraction Particle Size and Shape Analyzer

    0.01 - 3,500μm(Laser System)

    2 - 3,500μm(Image System)

  • Bettersizer ST

    Bettersizer ST

    One-stop Particle Size Analyzer

    Dispersion type: Wet

    Size range: 0.1 - 1,000µm

    Repeatability: ≤1% variation