What are the acceptable error ranges of repeatability?
2023-07-13WIKI
According to ISO 13320: 2020, six consecutive measurements should be carried out to satisfy the repeatability test. The average values of D10, D50, and D90 should be calculated, and the acceptable error ranges are:
♦ D10
Each measured value of D10 should not be deviated from the average value by more than ±2%.
♦ D50
Each measured value of D50 should not be deviated from the average value by more than ±1.5%.
♦ D90
Each measured value of D90 should not be deviated from the average value by more than ±3%.